Multiple Function Electrical Safety Analyzer
- Company Name:Ainuo Instrument Co.,Ltd.
- Membership:Free Member
- Member Since:2018. 10.24
- Country/Region:China
- City:Shandong
- Contact:Ada
- Related Keywords:Safety analyzer, Electrical Safety Analyzer
Ainuo Instrument Co.,Ltd.Contact us
Ainuo Instrument Co.,Ltd.
[China]
Main features
* Latest international standard:Comply with CCC,IEC,EN,VDE,BS,UL,JIS standard requirements,500VA WITHSTAND(AC 5KV/100mA),short-circuitoutput current exceeds 200mA,Ground resistance test current up to 64A,optional resistance or voltage mode.7 kinds of built-innetworking MD card for leakage contace current test,complying with multiple industrial standards.
★ Latest international standard: Comply with CCC, IEC, EN,VDE, BS, UL,JIS standard requirements. 500VA withstand (AC 5kV/100mA), short-circuit output current exceeds 200mA. Ground resistance test current up to 64A, optional resistance or voltage mode. 7 kinds of built-in networking MD card for leakage contact current test, complying with multiple industrial standards.
★ Intelligent test: 15 groups of test memories (group name editable), 8-step programming for each group. Product bar code reading and identification, automatically call the test group according to the model/specifications of the products tested. With U-disk to save test data timely.
★ Auto control: RS232/RS484/GPIB (four), bar code scanner interface, PLC interface, easily forming automation system.
★ Information management: optional electrical safety monitoring and control system software (ESRS), realizing test process and data management, high test efficiency, convenient for analysis and tracing.
★ Simplified operation: LCD display, shortcuts, Chinese/English.
Order information:
■AN9638H, 500VA(5000V/100mA) AC/DC withstand voltage, ground test (64A).
■AN9637H, 200VA(5000V/40mA) AC/DC withstand voltage, ground test (32A).
Spec
| AC Hipot test | ||
| Rated output capacity | AN9638H(F) | 500VA(5000V/100mA),short circuit current >200mA |
| AN9637H(F) | 200VA(5000V/40mA),short circuit current >200mA | |
| Output voltage setting | Range,Accuracy | 100~5000V,±(2%×setting+5counts) |
| Output frequency setting | Range,Accuracy | 50Hz or 60Hz,±0.1%×setting |
| Alarm current limits setting | Upper-limit range,Accuracy | 0.10~100.00mA(or 40.00mA),±(2%×setting+5counts) |
| Lower-limit range,Accuracy | 0.000~9.999mA,±(2%×setting+5counts) | |
| Time setting | Range | 0.5~999.9s(0 is infinite) |
| Ramp up/down time range | 0.1~999.9s | |
| Accuracy | ±(0.2%×setting+1count) | |
| Measurement | Voltage range,Accuracy | 0.01~5.00kV,±(1.5%×reading+1count) |
| Frequency range,Accuracy | 50Hz or 60Hz,±0.1%×reading | |
| Current range,Accuracy | 0.10~100.00mA(or 40.00mA),±(2%×reading+2counts) | |
| Time range,Accuracy | 0.1~999.9s,±(0.2%×reading+1count) | |
| Output voltage waveform, distortion, regulation | Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load) | |
| Start voltage setting | (0%~50%)×output voltage setting | |
| Current compensation setting | 0.000~100.00mA(or 40.00mA),auto、manual | |
| Arc detection | 1~9 level,9 level is optimally sensitive,0 is OFF | |
| DC Hipot test | ||
| Output voltage setting | Range,Accuracy | 100~6000VDC,±(2%×setting+5counts) |
| Alarm current limits setting | Upper-limit,Accuracy | 0~10000μA,±(2%×setting+5counts) |
| lower-limit,Accuracy | 0.0~999.9μA,±(2%×setting+5counts) | |
| Time setting | Range | 0.5~999.9s(0 is infinite) |
| Ramp up time range | 0.1~999.9s | |
| Ramp down time range | 0.3~999.9s(0 is OFF) | |
| Accuracy | ±(0.2%×setting+1 count) | |
| Measurement | Voltage range,Accuracy | 0.01~6.00VDC,±(1.5%×reading+1count) |
| Frequency range,Accuracy | 0.0~10000uA,±(2%×reading+5count) | |
| Current range,Accuracy | 0.1~999.9s,±(0.2%×reading+1count) | |
| Rated output capacity | 60VA(6000VDC/10mA) | |
| Output voltage ripple, regulation | <5%(pure resistance load,6000V/10mA),<(2%×setting+5V)(from empty load to full load) | |
| Compensation current Setting | 0 ~200.0μA,auto, manual | |
| Start voltage setting | (0%~50%)× setting output voltage | |
| Charging current lower limit | (0~3.500)μA,auto, manual | |
| Discharge time | ≤200ms | |
| Arc detection | 1~9 level,9 level is optimally sensitive,0 is OFF | |
| Insulation Resistance test | ||
| Output voltage setting | Range,Accuracy | 100~2500Vdc,±(2%×setting+5counts) |
| Alarm resistance setting | Range | 1~9999MΩ |
| Accuracy | 100-499V,1~1000MΩ,±(8%×setting+2counts) 1000~2000MΩ,±(12%×setting+2counts) 500-2500V,1~199MΩ,±(2%×setting+2counts), 200~999MΩ, ± (5%×setting+2counts) 1000~9999MΩ, ± (15%×setting+2counts) |
|
| Time setting | Ramp up time range | 0.1~999.9s |
| Adjustment delay time | 0,0.5~999.9s(0 is infinite) | |
| Test time range | 0,0.5~999.9s(0 is infinite) | |
| Ramp down time range | 0.1~999.9s | |
| Accuracy | ±(0.1%×setting+0.1s) | |
| Measurement | Voltage range, accuracy | 100~2500Vdc,±(2%×reading+5counts) |
| Resistance range, accuracy | 1~9999MΩ; 100-499V,1~1000MΩ,±(8%×reading+2counts) 1000~2000MΩ,±(12%×reading+2counts) 500-2500V,1~199MΩ,±(2%×reading+2counts), 200~999MΩ, ± (5%×reading+2counts) 1000~9999MΩ, ± (15%×reading+2counts) |
|
| Time range,accuracy | 0.1~999.9s,±(0.2%×reading+1counts) | |
| Charging current lower limit | 0~3.500μA,auto, manual | |
| Discharge time | ≤200ms | |
| Ground Bond test | ||
| Rated output | AN9638H(F) | Max output current is 64A,max testing resistance 600mΩ,open-circuit voltage <12V |
| AN9637H(F) | Max output current is 32A,max testing resistance 600mΩ,open-circuit voltage <12V | |
| Output current setting | Range,Accuracy | 2.0~32.0A(or 64A),±(2%×setting+5counts) |
| Output voltage setting | Range,Accuracy | 3.0~10.0V ,±(2%×setting+5counts) |
| Output frequency setting | Range,Accuracy | Sine wave,50Hz or 60Hz,±0.1%×setting |
| Alarm limits setting | Resistance upper/lower limits | 2.0~10.6A:0.1~600mΩ; 10.7~32.0A(or 64A):0.1~R mΩ,among R=(6400 / set current value)mΩ,when current range changs,Tester for automatic calculation range scope biggest |
| Tolerance | ±(2%×setting+5counts) | |
| Compensation resistance setting | Range | 0~100 mΩ,automatic measurement,ON/OFF |
| Test time setting | Range,Accuracy | 0.5~999.9s(0 stand for time infinite length);±(0.2%×setting+1count) |
| Measurement | Current range,accuracy | 2.0~32.0A(or 64A),±(2%×reading+2counts) |
| Voltage range,accuracy | 3.0~10.0V ,±(2%×reading+5counts) | |
| Resistance range, accuracy | 10.0~99.9~100~600mΩ;±(2%×reading+5counts) (64A current calculated method is same) |
|
| Time range,accuracy | 0.5~999.9s(0 stand for time infinite);±(0.2%×setting+1count) | |
| Extended function | ||
| Communication function | Address setting | 0~999 |
| Baud rate setting | 300,1200,2400,4800,9600,192000 | |
| Verification mode setting | None、odd、even | |
| Bar code scanning | Bar code scanning function | ON/OFF |
| Bar code length setting | 1~30 | |
| Model identification number | Start bit select 1~30,max length 16 Bit | |
| Function allocation explain | Hardware interface is optional | |
| print function | printer type setting | stylus printer miniprinter |
Other products from this supplier
Related products from premium suppliers on EC21
Browse: Manufacturer Directory Premium Suppliers Site Map
About EC21 Contact Us Terms & Conditions Report Item Online Trading Risks Product Listing Policy
Copyright (c)1997-2026 EC21 Inc. All Rights Reserved. EC21 in KoreanChinese
Business Registry Number: 120-86-03931 | Mall order sales Registration Number: 강남-5838