Ainuo Instrument Co.,Ltd. Company Logo Ainuo Instrument Co.,Ltd.

Precision Electrical Safety Comprehensive Tester

Contact Now* Send an Inquiry to this supplier.
or
Start Order * Name your price

Contact us

Ainuo Instrument Co.,Ltd.
[China]

Address
Ainuo Instrument Co.,Ltd. Qingdao Shandong
Phone
86-0532-83995118
Contact name
Ada

Quick Information

  • Brand Name: Ainuo
  • Place of Origin: China
  • Model Number : AN9651TH(F)
  • Power : Electronic
  • Usage : Auto Testing Machine

Description

Sigle-phase/Up to 7 functions
Double load for paralleled testing; 7 functions in 1 electrical safety and properties’ test; double channel display together; electric ARC, load open-circuit and short-circuit test function.
Applicable to such small home appliances high-speed product lines.

Technical characteristics:
◆ Compliant with GB4706.1-2005, IEC60335-1:2004, UL60335-1:2002, GB4943.1-2001, standard requirements.
◆ Three phase,single phase load compatible; Three phase rated load capacity 20kVA, can custom 60kVA.
◆ Seven functions into one device: ground bond, insulation resistance, AC hipot, DC hipot, leakage current, power test and low voltage start test.
◆ 15-group / 8-step test condition, intelligent judgment, auto compensation, auto protection;High speed test, the fastest speed can be 0.5s.
◆ Voltage ramp-up and ramp-down, discharge function;Dynamic and static leakage current, hipot test.
◆ Auto compensation, auto protection, auto memory, auto judgment, voice and light alarm.
◆ Serial communication, print, foot pedal interface, outer alarm light, PLC, network card interface.
◆ For safety test, the precision is 2%, for power test, the precision is 5%;For DUT, can support both grounding and suspending.
◆ Function preview, single screen setting, no need flip, simple operation.
◆ Support both 50Hz and 60Hz frequency, Chinese and English operation interface.

Model AN9651TH (F)(Three Phase/single phase7 in1)
Ground Bond Test
Empty Load Output Voltage <12V
Output Current Setting Range AC3A~30A 50Hz/60Hz,1A/step
Output Current Setting Accuracy ±(2% of setting+0.2A)
Upper Limit of Resistance 3A≤output current≤10A, 10~600mΩ; 10A
Ground Bond Resolution/ Measurement Range/ Accuracy 1mΩ, 10~600mΩ, ±(2% of reading +2mΩ)
Compensation Resistance Measurement Range/Accuracy 0~200mΩ 1mΩ
Test Time Setting Range 0.5~999.9s 0.1s/step
Insulation Resistance Test
Rated Output Load/Current 1VA (1000V/1mA) 1mA
Output Voltage Range/Adjustable DC:500V/1000V <1% (Empty load to full load)
Output Voltage Ripple Pure resistance load, 1000V/1mA, <3%
Insulation Resistance Range 1.0~2000MΩ
Insulation Resistance Lower Limit Setting Range/ Resolution 1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step
Insulation Resistance Accuracy 1~200 MΩ: ±(2% of reading+2counts) , 200~2000 MΩ: ±(5% of reading+5counts)
Ramp-up Time Range/Resolution 0.1~999.9s 0.1s
Delay Judgment Time 0.3~10.0s (Count from ramp-up )
Test Time Setting Range 0.5~999.9s 0.1s/step
AC Hipot Test
Rated Output Load 500VA (AC:5000V/100mA)
Output Voltage Range /Frequency/Adjustable Rate 500~5000V, 50Hz/60Hz, <2%(Empty load to full load)
Output Voltage Setting Range/Accuracy Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)
Leakage Current Preset Alarm Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step
Compensation Current Setting Range/ Resolution 0.00 ~5.00mA, 0.01mA
Output Initial Voltage Range (0%~99%) of test voltage, from 0V to initial voltage need 200ms
Leakage Current Measurement Range/Resolution/Accuracy 0.10~99.99 mA, 0.01mA, ±(2% of reading +2counts)
Test Time/Ramp-Up/Ramp-Down 0.5~999.9s, 0.1~999.9s, 0.0~999.9s(After test pass)
DC Hipot Test
Rated Output Load 60VA (AC:6000V/10mA)
Output Voltage Range /Frequency/Adjustable Rate 500~6000V, <2%(Empty load to full load)
Output Voltage Setting Range/Accuracy Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)
Leakage Current Preset Alarm Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA
Compensation Current Setting Range/ Resolution 0 ~5000μA, 1μA
Output Initial Voltage Range (0~99%) of test voltage, from 0V to initial voltage need 200ms
Leakage Current Measurement Range/Resolution/Accuracy 50~9999μA , 1μA, ±(2% of reading +2counts)
Test Time/Ramp-Up/Ramp-Down 0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from ramp-up)
Leakage Current Test
Output Voltage Range 1.06 times of input voltage (output by isolated transformer)
Leakage Current Range/Resolution Upper limit: 50~9999μA 1μA
Compensation Current Setting Range/ Resolution 0~500μA 1μA
Test Time Range/Resolution 1.0 ~999.9s 0.1s
Leakage Type dynamic, static
Three Phase Dynamic Mode A mode: ABC-AB-BC-CA (standard)
B mode: ABC-AB-ABC-BC-ABC-CA (customized)
Output Voltage Measurement Range/Accuracy AC 50~500.0V 0.1V, ±(0.5% of reading+2counts)
Leakage Current Measurement Range/Accuracy 50~9999μA 1μA 
DC~10kHz: ±(2% of reading+5counts) 10kHz~1MHz: ±(5% of reading+5counts)
Leakage Current Accuracy 50.0~2000μA is ±(5% of reading);2000μA or above is ±(2.5% of reading)
Leakage Power 6kVA ,isolated transformer capacity is designed according to the rated voltage 220V
Time Setting Range 1~300s
Measurement Mode dynamic, static
Power Test
Output Voltage Range 1.00 times of input voltage (output by isolated transformer)
Power Setting Range/Resolution Upper limit, Single phase:20~6000W Three phase:60~20000W
Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step
Test Time Range/Resolution 0.5 ~999.9s 0.1s
Voltage Measurement Range and Accuracy AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
Active Power Measurement Range and Accuracy 20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW 
45~65Hz: ±(0.5% of reading +2counts)
Power Factor Range and Accuracy 0.100~1.000, ±(2% of reading +2counts)
Low Voltage Start Test
Output Voltage Range 0.85 times of input voltage (output by isolated transformer)
Test Time Range/Resolution 0.5 ~999.9s 0.1s
Current Setting Range/Resolution Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A
Voltage Measurement Range and Accuracy AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
Display Mode 10.4 Inch, 640×480, TFT LCD display
Isolated Transformer Capacity Three Phase:20kVA, Single Phase: 6kVA
Over Current Protection Power, Start, Leakage Current, Phase Current35A, over 5s protection
DUT Type Setting Single Phase, Three phase Three line, Three phase Four line
Grounding Mode Load Grounding, Load Floating
DUT Error Processing Voice and light alarm
Outer Alarm Light Red alarm light
PLC Output Interface Each test function have one
Interface Function RS232, Outer Printer, Remote
Optional Interface RS485, GPIB, Internet Card, Inner Printer
Remote control Interface Foot Pedal, three-color alarm light(on-off signal output, PRINTER, 97 COM interface, Mini-Printer function optional)
Test Group 15 groups, 8 steps / group
Buzzer Alarm Voice Can turn on ,Can cut
 
 
 


Packaging & Delivery

  • Supply Ability : 2000

Product Image

  • Precision Electrical Safety Comprehensive Tester  image

Send an Inquiry to this supplier

Send an Inquiry to this supplier
* From
  To Ada
Ainuo Instrument Co.,Ltd.
* Buying Product
- Please enter your specific buying item.
e.g. 42 inch LCD TV, leather executive chair
Category : Testing Equipment
* Message

Use English only Max. 2000 characters. (Min. 20)

Send
EC21.com does not guarantee the validity of product information or the credentials of sellers.

Browse: Manufacturer Directory Premium Suppliers Site Map

About EC21 Contact Us Terms & Conditions Report Item Online Trading Risks Product Listing Policy

Copyright (c)1997-2026 EC21 Inc. All Rights Reserved. EC21 in KoreanChinese

Business Registry Number: 120-86-03931 | Mall order sales Registration Number: 강남-5838