| AC Hipot test |
|
| Booster rated output capacity |
500VA(5000V/100mA) |
|
| Booster maximum short-circuit current |
>200mA |
|
| Output waveform |
Sine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load) |
|
| |
| Output voltage setting |
Range |
500~5000V |
|
| Resolution |
5V |
|
| Accuracy |
±(3%×setting+10V) |
|
| Frequency setting |
Range |
50Hz/60Hz |
|
| Accuracy |
±0.1%×setting |
|
| Alarming limits setting |
Current upper-limit |
0.10~99.99mA |
|
| Current lower-limit |
0.00~10.00mA |
|
| Compensation current setting |
Range |
0.00~10.00mA, automatic testing, compensation can be set to ON/OFF |
|
| Time setting |
Range |
1~999s |
|
| Ramp up time range |
1~100s |
|
| Accuracy |
±(0.1%×setting+1count) |
|
| Breakdown current alarm |
Range |
0.10~99.99mA |
|
| Accuracy |
±(3%×reading+3counts) |
|
| Insulation Resistance test |
|
| Rated output capacity |
2VA(1000V/2mA) |
|
| Output voltage regulation |
<3%(From no load to full load) |
|
| Output voltage ripple |
<1% |
|
| Output voltage setting |
Range |
DC 500Vor1000V |
|
| Accuracy |
±(3%×setting+3counts) |
|
| Alarming limits setting |
Resistance upper-limit range |
0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjudgment for upper limit) |
|
| Resistance lower-limit range |
0.3~99.9 MΩ, 100~2000MΩ |
|
| Time setting |
Range |
1~999s |
|
| Accuracy |
±(0.1%×setting+1count) |
|
| Insulation resistance measurement |
Range |
0.0~99.9 MΩ, 100~2000 MΩ |
|
| Accuracy |
≤200MΩ: ±(3%×reading+3counts) |
|
| >200MΩ:±(8%×reading+8counts) |
|
| Ground Bond Resistance test |
|
| No load output voltage |
<12V |
|
| Output current setting |
Range |
3~30A |
|
| Accuracy |
±(3%×setting+0.2A) |
|
| Current range instruction |
Optional: 3 ~ 64A test current |
|
| Current frequency setting |
Range |
Sine waveform 50Hz/60Hz |
|
| Accuracy |
±0.1%×setting |
|
| Test time setting |
Range |
1~999s |
|
| Accuracy |
±(0.1%×setting+1count) |
|
| Result display mode |
Ground resistance or voltage |
|
| Ground resistance display mode |
|
| Alarming limits setting |
Resistance upper-limit range |
3A≤output current≤10A:10~600 mΩ |
|
| |
| |
| Resistance lower-limit range |
0~100 mΩ |
|
| Compensation resistance setting |
Range |
0~200 mΩ, automatic testing, compensation can be set to ON/OFF |
|
| Ground bond resistance measurement |
Range |
10~600 mΩ |
|
| Accuracy |
±(3%×reading+ 5counts) |
|
| Ground voltage display mode |
|
| Alarming limits setting |
Voltage upper-limit range |
30~7500 mV |
|
| Voltage lower-limit range |
0~3000 mV |
|
| Compensation voltage setting |
Range |
0~1000 mV, automatic testing, compensation can be set to ON/OFF |
|
| Ground bond resistance measurement |
Range |
10~7500 mV |
|
| Accuracy |
±(3%×reading+ 50mV) |
|
| Leakage Current test |
|
| Leakage current test type setting |
Single phase load, static/dynamic leakage current |
|
| MD network selection |
Standard IEC 60335-1 figure 4 network;other test network optional |
|
| Alarming limits Setting |
Current upper-limit range |
0.050~9.999mA |
|
| Current lower-limit range |
0.000~5.000mA |
|
| Leakage current compensation setting |
Range |
0.000~1.000mA, automatic testing, compensation can be set to ON/OFF |
|
| Time test setting |
Range |
1~999s |
|
| Accuracy |
±(0.1%×setting+1count) |
|
| Leakage current measurement |
Range |
0.050~9.999mA |
|
| accuracy |
DC~10kHz:±(3%×reading+0.010mA) |
|
| |
| Output voltage measurement |
Range |
60~280V(phase voltage) |
|
| accuracy |
±(0.5%×reading+2counts) |
|
| Run Power Only forAN9651S(F)/AN9651F(F)/AN9651B(F) |
|
| Alarming limits setting |
Power upper-limit range |
6~6000W |
|
| Power lower-limit range |
0~6000W |
|
| Test time setting |
Range |
1~999s |
|
| Accuracy |
±(0.1%×setting+1count) |
|
| Voltage measurement |
Range |
60~280V(phase voltage) |
|
| Accuracy |
±(0.5%×reading+2counts) |
|
| Current measurement |
Range |
0.030~3.999A, 4.00~25.00A |
|
| Accuracy |
±(0.5%×reading+2counts) |
|
| Active power measurement |
Range |
30.0~199.9W:±(0.5%×reading+5W) |
|
| Accuracy |
200~6000W(or2000W):±(0.5%×reading+30W) |
|